Detailed explanation of the working principle of ozone gas analyzer

Views :
Update time : 2025-02-28

Detailed explanation of the working principle of ozone gas analyzer

Ozone gas analyzer is a precision instrument for detecting and measuring ozone (O) concentration in the environment, which is widely used in industry, environmental monitoring, water treatment, food processing and other fields. Understanding its working principle helps to better select, use and maintain the equipment. This paper will analyze the working principle of ozone gas analyzer in detail.

202502285471.jpg

1. The core principle of ozone gas analyzer

The ozone gas analyzer is mainly based on the ultraviolet absorption method. The core principle is to use the absorption properties of ozone molecules to specific wavelengths of ultraviolet light. Ozone has a strong absorption peak at a wavelength of 254 nanometers (nm), and the concentration of ozone can be calculated by measuring the change in light intensity before and after ultraviolet light passes through a gas sample.



2. The specific process of ultraviolet absorption method

- Light source emission: The instrument's built-in ultraviolet lamp emits ultraviolet light with a wavelength of 254 nm.

- Optical path transmission: ultraviolet light is transmitted through the gas chamber (containing the gas sample to be measured).

- Light intensity detection: After the ultraviolet light passes through the gas chamber, the photoelectric sensor detects the remaining light intensity.

- Concentration calculation: According to the Lambert-Beer Law, the concentration of ozone is calculated by comparing the intensity of the incident light and the intensity of the transmitted light.


3. Application of Lambert-Beer law

The Lambert-Beer law is the basis of the ultraviolet absorption method, and its formula is:


A = ϵ ⋅ ⋅ l c

Among them:


- (A) is absorbance;

- (ϵ) is the molar absorption coefficient of ozone;

- (c) Ozone concentration;

- (l) is the optical path length (air chamber length).

Ozone concentration (c) can be calculated by measuring absorbance (A), combined with the known (ϵ) and (l).


4. Key components of ozone gas analyzer

- UV light source: provides stable 254 nm UV light.

- Gas chamber: used to hold the gas sample to be measured, the length of the optical path affects the measurement accuracy.

- Photoelectric sensor: detects the transmitted light intensity and converts it into an electrical signal.

- Signal processing unit: The electrical signal is processed, the ozone concentration is calculated and the result is displayed.


5. Advantages of ozone gas analyzer

- High accuracy: UV absorption method has extremely high measurement accuracy and sensitivity.

- Real-time monitoring: continuous on-line monitoring of ozone concentration.

- Strong anti-interference ability: not affected by other gas components, reliable measurement results.

- High stability: The instrument is stable and suitable for long-term monitoring.


6. Application Scenarios

- Industrial ozone generator: Monitor the output concentration of the ozone generator.

- Water treatment: Test the amount of ozone added in the water treatment process.

- Environmental monitoring: Monitoring ozone concentrations in the atmosphere and assessing air quality.

- Food processing: for the control of ozone concentration during food disinfection.


7. Use precautions

- Periodic calibration: To ensure the accuracy of measurement results.

- Maintenance of light source and air chamber: avoid contamination or damage affecting measurement accuracy.

- Avoid high temperature and humidity environments: extreme environments may affect instrument performance.

2024122630025465.jpg

Conclusion

Ozone gas analyzer is based on the working principle of ultraviolet absorption method. It has the characteristics of high precision, real-time monitoring and strong anti-interference ability, and is an important tool for ozone concentration detection. By understanding its working principle, users can better choose and use the equipment to ensure the accuracy and reliability of the measurement results. If you need more information about ozone gas analyzers, please contact us!


Related News
Read More >>
臭氧检测仪测量结果不准确可能原因 臭氧检测仪测量结果不准确可能原因
03 .20.2025
臭氧检测仪测量结果不准确可能原因问题描述:测量值与实际值偏差较大,数据不稳定。可能原因:传感器老化或损坏。气室污染或光路受阻。校准不准确或未定期校准。环境温度、
Detailed explanation of the wo Detailed explanation of the wo
02 .28.2025
Detailed explanation of the working principle of ozone gas analyzerOzone gas ana
臭氧检测仪国家标准有哪些 臭氧检测仪国家标准有哪些
02 .20.2025
臭氧检测仪国家标准有哪些随着环保政策趋严和健康意识提升,臭氧检测仪在工业安全、环境监测及家庭场景的应用需求激增。为保障设备精准性与合规性,国家已更新多项国家标准
紫外吸收气体分析仪的结构与工作原理 紫外吸收气体分析仪的结构与工作原理
11 .26.2024
紫外吸收气体分析仪的结构与工作原理紫外吸收法是由于臭氧(O3)吸收特定波长的紫外线辐射,因此该方法可用于测量O3气体。由于O3气体在紫外区域的吸收波段为253.